Required Material
Aquadibidestilata
Procedure
Running 60 samples (Aquadibidestilata) for 12 Wavelength
(340, 380, 405, 450, 505, 546, 570, 600, 660, 700, 750 and 800 nm)
Requirement
Expected Result : (-20) - (+20)
Result
| Wave Length |
Number of Run |
Result |
Status |
Wave Length |
Number of Run |
Result |
Status |
| 340 nm |
60 |
1 |
PASS |
570 nm |
60 |
0 |
PASS |
| 380 nm |
60 |
1 |
PASS |
600 nm |
60 |
0 |
PASS |
| 405 nm |
60 |
0 |
PASS |
660 nm |
60 |
0 |
PASS |
| 450 nm |
60 |
0 |
PASS |
700 nm |
60 |
0 |
PASS |
| 480 nm |
60 |
0 |
PASS |
750 nm |
60 |
0 |
PASS |
| 505 nm |
60 |
0 |
PASS |
800 nm |
60 |
0 |
PASS |
| 546 nm |
60 |
0 |
PASS |
|
|
|
|
Conclusion
Instrument TMS 30i have been already calibrated.
The Result of Calibration is good.