Nomor: FO.III.12/04.00/2019

Certificate of Instrument Calibration

Customer Data

Laboratory / Hospital : Trisensa Diagnostic Center
Address : Jl. Taman AIS Yasin No.5, Embong Kaliasin, Kec. Genteng, Surabaya 60271
Instrument : Clinical Chemistry Analyzer TMS 50i Superior
Serial Number : 5303150816
Date of Calibration : January 09, 2026
Valid Until : January 09, 2027

Required Material

Aquadibidestilata

Procedure

Running 60 samples (Aquadibidestilata) for 12 Wavelength
(340, 380, 405, 450, 505, 546, 570, 600, 660, 700, 750 and 800 nm)

Requirement

Expected Result : (-20) - (+20)

Result

Wave Length Number of Run Result Status Wave Length Number of Run Result Status
340 nm 60 -2 PASS 570 nm 60 -1 PASS
380 nm 60 -1 PASS 600 nm 60 0 PASS
405 nm 60 0 PASS 660 nm 60 -2 PASS
450 nm 60 -1 PASS 700 nm 60 -2 PASS
505 nm 60 -3 PASS 750 nm 60 -1 PASS
546 nm 60 -1 PASS 800 nm 60 -2 PASS

Conclusion

Instrument TMS 24i Premium have been already calibrated. The Result of Calibration is good.