Nomor: FO.III.12/04.00/2019

Certificate of Instrument Calibration

Customer Data

Laboratory / Hospital : Trisensa Diagnostic Center
Address : Jl. Taman AIS Yasin No.5, Embong Kaliasin, Kec. Genteng, Surabaya 60271
Instrument : Clinical Chemistry Analyzer TMS 50i Superior
Serial Number : 5303150816
Date of Calibration : January 09, 2026
Valid Until : January 09, 2027

Required Material

Standart Solution 1
Standart Solution 2

Procedure

Running 60 samples (Aquadibidestilata) for 12 Wavelength

Requirement

Expected Result :
Slope Na : (+40) - (+72) mV/dec
Slope K : (+40) - (+67) mV/dec
Slope Cl : (-25) - (-62) mV/dec

Result

Type of Calibration Slope Na Status Slope K Status Slope Cl Status
Auto Calibration (+46.4) PASS (+52.2) PASS (-41.0) PASS
Manual Calibration (+46.6) PASS (+51.7) PASS (-42.0) PASS
Slope Wash Calibration (+44.5) PASS (+50.9) PASS (-40.9) PASS

Conclusion

Instrument JOKOH EX-D have been already calibrated. The Result of Calibration is good.