Nomor: FO.III.12/04.00/2019

Certificate of Instrument Calibration

Customer Data

Laboratory / Hospital : Trisensa Diagnostic Center
Address : Jl. Taman AIS Yasin No.5, Embong Kaliasin, Kec. Genteng, Surabaya 60271
Instrument : Clinical Chemistry Analyzer TMS 50i Superior
Serial Number : 5303150816
Date of Calibration : January 09, 2026
Valid Until : January 09, 2027

Required Material

Aquadibidestilata

Procedure

Running 60 samples (Aquadibidestilata) for 12 Wavelength
(340, 380, 405, 450, 505, 546, 570, 600, 660, 700, 750 and 800 nm)

Requirement

Expected Result : (-20) - (+20)

Result

Parameters Value Range Unit Result
Incubation Module 37.0 36.7 - 37.3 °C Pass
Reagent Compartment 5.0 2.0 - 8.0 °C Pass
Power 24V Voltage 23.94 20 - 30 Volt Pass
Power 12 V Voltage 11.87 11.0 - 14.0 Volt Pass
Vacuum Pressure Check -23.00 -24.00 - -34.00 KPa Pass
DCF Diagnosis 0.96 < 1.5% % Pass
Substrate Background 48 SD < 200 N/A Pass
System Repeatability 0.12% CV ≤ 1.5% % Pass
Dark Current Count 80.2 0 - 350 N/A Pass

Conclusion

Instrument Chemiluminescence Immunoassay System Mindray CL-900i have been already calibrated. The result of calibration is good.