Required Material
Aquadibidestilata
Procedure
Running 60 samples (Aquadibidestilata) for 12 Wavelength
(340, 380, 405, 450, 505, 546, 570, 600, 660, 700, 750 and 800 nm)
Requirement
Expected Result : (-20) - (+20)
Result
| Parameters |
Value |
Range |
Unit |
Result |
| Incubation Module |
37.0 |
36.7 - 37.3 |
°C |
Pass |
| Reagent Compartment |
5.0 |
2.0 - 8.0 |
°C |
Pass |
| Power 24V Voltage |
23.94 |
20 - 30 |
Volt |
Pass |
| Power 12 V Voltage |
11.87 |
11.0 - 14.0 |
Volt |
Pass |
| Vacuum Pressure Check |
-23.00 |
-24.00 - -34.00 |
KPa |
Pass |
| DCF Diagnosis |
0.96 |
< 1.5% | %
|
Pass |
| Substrate Background |
48 |
SD < 200 | N/A |
Pass |
| System Repeatability |
0.12% |
CV ≤ 1.5% |
% |
Pass |
| Dark Current Count |
80.2 |
0 - 350 |
N/A |
Pass |
Conclusion
Instrument Chemiluminescence Immunoassay System Mindray CL-900i have been already calibrated.
The result of calibration is good.