Nomor: FO.III.12/04.00/2019
Certificate of Instrument Calibration
Customer Data
|
|
|
Laboratory / Hospital
|
: Trisensa Diagnostic Center
|
|
Address
|
: Jl. Taman AIS Yasin No.5, Embong Kaliasin, Kec. Genteng, Surabaya 60271
|
|
Instrument
|
: Clinical Chemistry Analyzer TMS 50i Superior
|
|
Serial Number
|
: 5303150816
|
|
Date of Calibration
|
: January 09, 2026
|
|
Valid Until
|
: January 09, 2027
|
Required Material
-
Procedure
Running 60 samples (Aquadibidestilata) for 12 Wavelength
Requirement
Expected Result : (-20) - (+20)
Result
| Parameters |
Value |
Range |
Unit |
Result |
| Reaction Carousel Temperature |
37.0 |
36.7 - 37.3 |
°C |
Pass |
| Cuvette Cleaning Fluid Temperature |
44.0 |
39.0 - 49.0 |
°C |
Pass |
| Cuvette Wash Solution Temperature |
44.2 |
39.0 - 49.00 |
°C |
Pass |
| Cuvette Check |
No Yellow Cuvette |
|
N/A |
Pass |
| Photometer Check |
7997 |
5000-8000 |
λ |
Pass |
| Bar Code Stability Check |
Read |
Read / No Read |
N/A |
Pass |
Conclusion
Instrument BS-430 Chemistry Analyzer has been already calibrated.
The result of calibration is good.