Nomor: FO.III.12/04.00/2019

Certificate of Instrument Calibration

Customer Data

Laboratory / Hospital : Trisensa Diagnostic Center
Address : Jl. Taman AIS Yasin No.5, Embong Kaliasin, Kec. Genteng, Surabaya 60271
Instrument : Clinical Chemistry Analyzer TMS 50i Superior
Serial Number : 5303150816
Date of Calibration : January 09, 2026
Valid Until : January 09, 2027

Required Material

-

Procedure

Running 60 samples (Aquadibidestilata) for 12 Wavelength

Requirement

Expected Result : (-20) - (+20)

Result

Parameters Value Range Unit Result
Reaction Carousel Temperature 37.0 36.7 - 37.3 °C Pass
Cuvette Cleaning Fluid Temperature 44.0 39.0 - 49.0 °C Pass
Cuvette Wash Solution Temperature 44.2 39.0 - 49.00 °C Pass
Cuvette Check No Yellow Cuvette N/A Pass
Photometer Check 7997 5000-8000 λ Pass
Bar Code Stability Check Read Read / No Read N/A Pass

Conclusion

Instrument BS-430 Chemistry Analyzer has been already calibrated. The result of calibration is good.