Nomor: FO.III.12/04.00/2019
Certificate of Instrument Calibration
Customer Data
|
|
|
Laboratory / Hospital
|
: Trisensa Diagnostic Center
|
|
Address
|
: Jl. Taman AIS Yasin No.5, Embong Kaliasin, Kec. Genteng, Surabaya 60271
|
|
Instrument
|
: Clinical Chemistry Analyzer TMS 50i Superior
|
|
Serial Number
|
: 5303150816
|
|
Date of Calibration
|
: January 09, 2026
|
|
Valid Until
|
: January 09, 2027
|
Required Material
|
SC Cal |
Procedure
|
Checking Instrument Status, Running Callibartion
|
Requirement
|
Deviation must be ≤ 2.0% ; 3 Impedance measurement must be OK; Measurement point must be within range
|
Results
Calibration Factor
| Parameters |
Reference |
Value |
Result |
| WBC |
75 - 125 |
95.20 |
Pass |
| RBC |
75 - 125 |
105.38 |
Pass |
| HGB |
75 - 125 |
101.86 |
Pass |
| MCV |
75 - 125 |
104.22 |
Pass |
| PCT |
75 - 125 |
105.58 |
Pass |
| RBC-O |
75 - 125 |
- |
- |
| PCT-O |
75 - 125 |
- |
- |
| PCT-H |
75 - 125 |
- |
- |
|
Optical
| Parameters |
Deviation |
Result |
| DIFF_FS |
0.1 |
Pass |
| DIFF_SS |
0.8 |
Pass |
| DIFF_FC |
0.5 |
Pass |
| DIFF_PMT |
N/A |
Pass |
| RET_FS |
- |
- |
| RET_SS |
- |
- |
| RET_FC |
- |
- |
| RET_PMT |
- |
- |
|
Impedance
| Parameters |
Target |
1 |
2 |
3 |
Result |
| MCV_G |
93.1 |
97.9 |
92.7 |
93.3 |
OK |
| HGB |
4.40 |
4.75 |
4.40 |
4.40 |
OK |
ESR Measurement Point
| Parameters |
Tested |
Reference |
Result |
| Measurement Point (µL) |
- |
135 - 180 |
- |
Conclusion
Instrument Auto Hematology Analyzer BC-760R has been already calibrated.
The result of calibration is good.