Nomor: FO.III.12/04.00/2019

Certificate of Instrument Calibration

Customer Data

Laboratory / Hospital : Trisensa Diagnostic Center
Address : Jl. Taman AIS Yasin No.5, Embong Kaliasin, Kec. Genteng, Surabaya 60271
Instrument : Clinical Chemistry Analyzer TMS 50i Superior
Serial Number : 5303150816
Date of Calibration : January 09, 2026
Valid Until : January 09, 2027

Required Material

SC Cal

Procedure

Checking Instrument Status, Running Callibartion

Requirement

Deviation must be ≤ 2.0% ; 3 Impedance measurement must be OK; Measurement point must be within

Results

Parameters Value Range Unit Result
Power +12V 11.79 11.00 - 14.00 V Pass
Power +24V 24.06 20.00 - 30.00 V Pass
Analog +12V 11.70 11.00 - 13.00 V Pass
Analog -12V -11.72 -14.00 - -9.00 V Pass
Digital +56V 56.05 47.00 - 63.00 V Pass
HGB Blank Voltage 4.21 3.85 - 4.85 V Pass
FS Blank Voltage 0.29 0.07 - 1.00 V Pass
Laser Diode Current 28 20 - 70 mA Pass
Diluent Temperature 27 10.00 - 40.00 °C Pass
Reagent preheating Temperature 42.2 39.5 - 44.0 °C Pass
Optical System Temperature 29.4 15.0 - 40.0 °C Pass
Liquid Pressure 92.0 70.00 - 110.0 kPa Pass
Vacuum -28.6 -35.0 - -26.0 kPa Pass
WBC 99.93 75 - 125 % Pass
RBC 101.49 75 - 125 % Pass
HGB 99.65 75 - 125 % Pass
MCV 99.94 75 - 125 % Pass
PLT 101.60 75 - 125 % Pass

Conclusion

Instrument Auto Hematology Analyzer BC-5140 has been already calibrated. The result of calibration is good.