Required Material
SC Cal
Procedure
Checking Instrument Status, Running Callibartion
Requirement
Deviation must be ≤ 2.0% ; 3 Impedance measurement must be OK; Measurement point must be within
Results
| Parameters |
Value |
Range |
Unit |
Result |
| Power +12V |
11.79 |
11.00 - 14.00 |
V |
Pass |
| Power +24V |
24.06 |
20.00 - 30.00 |
V |
Pass |
| Analog +12V |
11.70 |
11.00 - 13.00 |
V |
Pass |
| Analog -12V |
-11.72 |
-14.00 - -9.00 |
V |
Pass |
| Digital +56V |
56.05 |
47.00 - 63.00 |
V |
Pass |
| HGB Blank Voltage |
4.21 |
3.85 - 4.85 |
V |
Pass |
| FS Blank Voltage |
0.29 |
0.07 - 1.00 |
V |
Pass |
| Laser Diode Current |
28 |
20 - 70 |
mA |
Pass |
| Diluent Temperature |
27 |
10.00 - 40.00 |
°C |
Pass |
| Reagent preheating Temperature |
42.2 |
39.5 - 44.0 |
°C |
Pass |
| Optical System Temperature |
29.4 |
15.0 - 40.0 |
°C |
Pass |
| Liquid Pressure |
92.0 |
70.00 - 110.0 |
kPa |
Pass |
| Vacuum |
-28.6 |
-35.0 - -26.0 |
kPa |
Pass |
| WBC |
99.93 |
75 - 125 |
% |
Pass |
| RBC |
101.49 |
75 - 125 |
% |
Pass |
| HGB |
99.65 |
75 - 125 |
% |
Pass |
| MCV |
99.94 |
75 - 125 |
% |
Pass |
| PLT |
101.60 |
75 - 125 |
% |
Pass |
Conclusion
Instrument Auto Hematology Analyzer BC-5140 has been already calibrated. The result of calibration
is good.